Lab Equipment Lab Science Auto Chemistry Analyzer Model ME200


Reaction tray
37ºC±0.1ºC, real-time monitor. 

MixerProbe
Teflon coating to avoid cross contamination. 
Reagent tray
2ºC~8ºC cooling for 24 hours. 
Sample needle
Liquid level sensor function. Anti-collision function.Reagent volume real-time detection. 
Software
User-friendly software. 
Washing Probe
Independent 4-step washing system. 
Application:
Automatic medical washing disinfector provides standard and automatic cleaning and disinfection for most of medical equipments , so as to eliminate the possibility of infection in cleaning and disinfecting .
Features:
1.49 Sample positions.
2.56 Reagent positions.
3.120 Reaction cuvettes
4.200 tests per hour.
5.Probe with anti-collision function, liquid level dectection function.
Specifications:
Model  | ME-200  | |
Overall Performance  | Throughput  | 200Tests/Hour  | 
Analysis Method  | End-point, Fixed-time, Rate(Kinetic), Turbidimetry  | |
Certificates  | CE, FDA, ISO9001, ISO14001, ISO13485  | |
Sample & Reagent Unit  | Sample Positions  | 49 sample positions  | 
Reagent Positions  | 56 reagent positions  | |
Sample Volume  | 2~70μl  | |
Reagent Volume  | 20~350μl  | |
Probe  | Teflon coating, withanti-collision function, liquid leveldetectionfunction.  | |
Probewashing  | Automaticwashing interior and exterior  | |
ReagentCooling  | Refrigerated tray with independent switch  | |
Reaction System  | Temperature Control  | 37±0.1ºC, real-time monitor  | 
Cuvettes  | 120 reusable cuvettes,optical length 6mm  | |
MixerProbe  | Independent stirring  | |
Washing  | Automatic cuvettes washing  | |
STAT function  | YES  | |
Optical System  | Light Source  | 12V/30W halogen lamp  | 
Spectrophotometry  | Post-spectral spectrophotometry  | |
Wavelength  | 340,405,450,510,546,578,630,700nm  | |
Absorbance  | 0~3.0Abs  | |
Calibration&QC  | Calibration  | Linear: K factor, 1-point, 2-point and multipoint linear Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P,Logit-5P  | 
Quality Control  | Real-time QC, Westgard multi rule, Cumulative sum check, Twin Plot(2D)  | |
Data Management  | Software  | Windows 7/8/10, LIS system available  | 
LIS System  | Bi-direction, support HL7 protocol  | |
Interface  | LAN port access  | |
Printer  | External, multiple reporting mode available  | |
Working Conditions  | Power Supply  | AC220V±10%, 60/50Hz,, 110V±10%, 60Hz, 300W  | 
Temperature  | 10~30ºC  | |
WaterConsumption  | Deionized water:5L/H  | |
Humidity  | 30-80%  | |
Size & Weight  | External Size (W*D*H)  | 950*612*510mm  | 
Net Weight  | 75kg  | |
Package Size(W*D*H)  | 1118*728*1151mm  | |