Online Total Organic Carbon (TOC) and Conductivity Analyzers

Features:1. The CAN bus communication, a data terminal CAN be configured at the most eight detection unit, eight monitoring CAN detect at the same time.2. Implement for multi-point synchronous water supply system, continuous on-line detect, save cost, convenient operation, reduce the workload.3. 8 first, focus on display, monito

Featured Products

Features:
1. The CAN bus communication, a data terminal CAN be configured at the most eight detection unit, eight monitoring CAN detect at the same time.
2. Implement for multi-point synchronous water supply system, continuous on-line detect, save cost, convenient operation, reduce the workload.
3. 8 first, focus on display, monitor, display the TOC, conductivity and temperature at the same time.
4. Can be set arbitrary analysis time interval and data storage time interval.
5. Has automatic upper limit alarm output function, timely remind when abnormal data.
6. With uv lamp period countdown function, accurate calculation of uv lamp use time.
7. Data can be stored query, large data storage capabilities, data can be preserved more than 5 years.
8. With USB interface, can deduce the data by U disk
9. With functions of print output.
10. The swappable offline detection model for monitoring, calibration and verification.
11. Without adding acid reagent, antioxidants and any gas, without additional daily maintenance.
12. The operation is simple, fast and reliable, users don't need professional knowledge and special training.
13. Small volume, light weight, less energy consumption.
14. A waterproof design, can work under the condition of high humidity and temperature.
15. Compact cabinet design meets the requirement of harsh industrial environments.

Technical parameters:
TOC detection range: 0.0001 mg/L ~ 1.0000 mg/L
Conductivity detection range: 0.055 mu s/cm ~ 6.000 mu s/cm
The TOC precision: plus or minus 5%
The TOC detection limit: 0.0001 mg/L (conductivity < 1.0 mu s/cm)
Analysis of time: the analysis of the 200 s shortest time interval
Response time: within 15 min
Repeatability error: 3% or less
Zero drift: plus or minus 5%
Span drift: plus or minus 5%
The sample temperature: 1 ~ 95 °C
Ambient temperature: 10 ~ 60 °Ctemperature changes within + / - 5 °C / d
Power source: 22 v to 220 v +
Power frequency: 50 hz + 1 hz
Rated power: 100 w
The basic size: 32 cm * 19 cm * 21 cm

Online Total Organic Carbon (TOC) & Conductivity Analyzer


Factory View
Online Total Organic Carbon (TOC) & Conductivity AnalyzerOnline Total Organic Carbon (TOC) & Conductivity Analyzer
Online Total Organic Carbon (TOC) & Conductivity Analyzer
Online Total Organic Carbon (TOC) & Conductivity Analyzer

Contact us