Clinic Medical Automatic Chemistry Biochemistry Analyzer
Technical Specifications
System Functions
-Analysis Method: end-point, kinetic, two-point, double-reagent, double-wavelength, multi-standard etc, open to various reagents
Reaction System
- Reaction Cuvette: 60 reusable plastic units
- Optical Length of Cuvette: 6mm
- Reaction Volume: 180~400µl
-Max reaction Time: 10 minutes
- Reaction Temperature: 37±0.1ºC
Optical System
- Light Source: halogen-tungsten lamp
- Wavelength: 340-690nm, 9 wavelengths, precision ±1.5nm
- Absorbance Range: 0~5.0Abs
- Resolution: 0.0001Abs
- Rear spectrophotometry
Calibration
-Calibration: linear/nonlinear multi-points calibration
-Re-test: retest the sample automatically when the result is out of the linearity range or the sample is not sufficient
Working Condition
- Power Supply: ~100-240V, 50/60Hz, 300VA
- Temperature: 10-35ºC
- Humidity: ≤90%, no dew
Operation System
- Windows XP or Windows 7
Input and Output
- Input: RS-232 interface/computer
-Output: multi-format printout
Sample/Reagent Position
-18pcs sample position
-Reagent refrigerated function
-26pcs reagent position
Reaction Cuvettes
- 60 pcs reusable plastic cuvettes
- Automatic blank testing for cuvette
-Colorimetry system in the pattern
of light splitting behind cuvette
-Dismountable, easily replacement
Sample/Reagent/Mixing Probe
-Automatic liquid level detect
-Collision protection function
-Probe internal/external washing
- Independent mixing probe
7-Step Cleaning System
-Automatic 7-step washing
-Hi-tech washing station to ensure
accurate results
-With low water consumption: 1-2L/H
Intellective Software
-Automatic washing cuvette when startup & shut down
-Automatic detect reagent volume during every startup
- Reagent plate temperature detecting when startup
Optimum calibration cuvette
-Automatic Cuvette blank testing
-Automatic wash selected cuvettes
Dynamic and Real-time display of running status
-Real-time online monitoring running status of sample tray, reagent tray and cuvette
-Real-time online monitoring reagents residual volume
- Real-time online monitoring waste volume
Automatic dilute and re-test
-Automatic dilute and re-test when absorbance range is out of range
-Automatic dilute and re-test when linear limit is out of range
-Automatic dilute and re-test when linear range is out of range
-Freely set the auto dilute ratio and dilute water position
A revolutionary automatic Chemistry analyzer with
Optimal substitution of semi-auto Chemistry
analyzer
Shipping & Payment
For total weight over 100kgs, shippment by air or by sea is suggested.
Contact Us
Technical Specifications
System Functions
-Analysis Method: end-point, kinetic, two-point, double-reagent, double-wavelength, multi-standard etc, open to various reagents
Sample/Reagents Handling -Sample volume: 1.6-50µl, 0.1µl step -Sample Dilution: automatic/Manual pre-dilution, with dilution ratio up to 1:100 -Reagent Position: 26 pcs (Refrigerated function is optional ) -Reagent Volume: 10-500µl, 0.5µl step |
Reaction System
- Reaction Cuvette: 60 reusable plastic units
- Optical Length of Cuvette: 6mm
- Reaction Volume: 180~400µl
-Max reaction Time: 10 minutes
- Reaction Temperature: 37±0.1ºC
Optical System
- Light Source: halogen-tungsten lamp
- Wavelength: 340-690nm, 9 wavelengths, precision ±1.5nm
- Absorbance Range: 0~5.0Abs
- Resolution: 0.0001Abs
- Rear spectrophotometry
Calibration
-Calibration: linear/nonlinear multi-points calibration
-Re-test: retest the sample automatically when the result is out of the linearity range or the sample is not sufficient
Working Condition
- Power Supply: ~100-240V, 50/60Hz, 300VA
- Temperature: 10-35ºC
- Humidity: ≤90%, no dew
Operation System
- Windows XP or Windows 7
Input and Output
- Input: RS-232 interface/computer
-Output: multi-format printout
Sample/Reagent Position
-18pcs sample position
-Reagent refrigerated function
-26pcs reagent position
Reaction Cuvettes
- 60 pcs reusable plastic cuvettes
- Automatic blank testing for cuvette
-Colorimetry system in the pattern
of light splitting behind cuvette
-Dismountable, easily replacement
Sample/Reagent/Mixing Probe
-Automatic liquid level detect
-Collision protection function
-Probe internal/external washing
- Independent mixing probe
7-Step Cleaning System
-Automatic 7-step washing
-Hi-tech washing station to ensure
accurate results
-With low water consumption: 1-2L/H
Intellective Software
-Automatic washing cuvette when startup & shut down
-Automatic detect reagent volume during every startup
- Reagent plate temperature detecting when startup
Optimum calibration cuvette
-Automatic Cuvette blank testing
-Automatic wash selected cuvettes
Dynamic and Real-time display of running status
-Real-time online monitoring running status of sample tray, reagent tray and cuvette
-Real-time online monitoring reagents residual volume
- Real-time online monitoring waste volume
Automatic dilute and re-test
-Automatic dilute and re-test when absorbance range is out of range
-Automatic dilute and re-test when linear limit is out of range
-Automatic dilute and re-test when linear range is out of range
-Freely set the auto dilute ratio and dilute water position
A revolutionary automatic Chemistry analyzer with
- Compact design : 54.8cmX42.2cmX38.3cm
- Easy to carry: net weight ≤23kg
Optimal substitution of semi-auto Chemistry
analyzer
- Minimal labor required
- Independent mixing probe
- Utility of reagents decreased: 300ul/test
- Efficiency speed: 200 tests/hour
- User friendly : Smart and easy-operating software
- High Accuracy: Minimal operating /random errors
- High reputation for Clinic
Shipping & Payment
For total weight over 100kgs, shippment by air or by sea is suggested.
Contact Us