Features
Analysis System
Optical System
Reagent/Sample Tray
- Random access, discrete automatic biochemistry analyzer, insert emergency at any time
- High performance and reasonable price, It is the best economical choice from semi-auto upgraded to fully-auto analyzer
- Probe performance: Liquid level detection, all-round anti-collision protection and inventory checking, have automatic stop and alarm prompt for an obstacle
- Probe cleaning: Automatic washing for both interior and exterior
Analysis System
- Test Speed: constant 160tests/hour, up to 220tests/hour
- Test method: End-point, Fixed-time (two point method), kinetic
- Reaction tray: 81 UV cuvettes (disposable)
- Mini reaction volume: 150μl, save reagent cost
- Insert emergency at any time
- Support linear and non-linear calibration
- Heating method: solid constant temperature closed direct heating
- Temperature equilibrium time: less than 3minutes,
- Reaction Temperature: 37±0.2°C
- Temperature Fluctuation: ± 0.2°C
Optical System
- Integrated full closed optical system, after spectrometric method, no mechanical parts, not moldy, no maintenance cost
- 8 wavelengths, selectable among 340-800nm
- Absorption range: 0.00-4.00Abs
- Halogen-tungsten lamp has more than 2000hrs service time
- Center wavelength deviation: ±2nm, half wave width less than 12nm
- Stray light less than 0.3%
- Resolution: 0.0001Abs
Reagent/Sample Tray
- 60 reagent + 30 sample positions
- Advanced reagent tray refrigeration system, temperature: 4-12°C (optional function)
- Automatically detect reagent remain, real time display reagent remain
- Reagent bottles with 15ml,30ml,50ml,70ml are available.
- Support single/ dual reagents, reagents are open
- Real time display reagent tray status
- Automatically judge substrate depletion
Technical parameter: | |
Equipment type | Fully automatic discrete, STAT priority |
Analysis rate | Constant speed 160T/H (without ISE) |
Test principle | Colorimetry, turbidimetry |
Analysis method | End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration |
Sample Tray | 30 sample positions, random access of all samples |
Sample Volume | 2-50ul, 0.1ul increment |
Sample probe | Liquid level detection, the remaining detection, collision protection etc. |
Sample Containers | Sample cups, Micro tube, Original blood-collection tube, etc. |
Reagent Tray | 60 reagent positions |
Reagent Volume | R1:25-300ul, R2:10-150ul, 0.1ul increment |
Reagent Probe | Liquid level detection, the remaining detection, collision protection etc. |
Reagent Refrigeration | 4-15ºC, independent power supply, 24-hour non-stop refrigeration |
Reaction Tray | Rotating tray, 120 cuvettes |
Cuvette | eusable high quality cuvette, Optical diameter 6mm |
Reaction Volume | 180-500ul |
Incubation system | Bi-directional liquid-air device |
Power supply | AC 220V±22V, 50Hz±1Hz, power 1.5KVA |
Ambient Temperature | 15-30 ºC |
Humidity | 35-80% |
Water consumption | 3L/Hour |
Dimension | Bench top: 750mm*520mm*600mm |